Page 53
Appendix | Industrial Process Interface (Optional)
thermoIMAGER TIM
The industrial process interface offers the following inputs and outputs:
Designation Description max. range
1
/ status
A IN 1 / 2 Analog input 1 und 2 0 - 10 V
D IN 1 Digital input 24 V
AO1 / 2 / 3 Analog output 1, 2 and 3 0 - 10 V
Alarm output 1, 2 and 3 0 / 10 V
DO1 / 2 / 3 Relay output 1, 2 and 3
2
open/ closed (red LED on) / 0 ... 30 V, 400 mA
FS Fail-safe relay open/ closed (green LED on) / 0 ... 30 V, 400 mA
The process interface has an integrated fail-safe mode. This allows to control conditions like interruption of
cables, shut-down of the software etc. and to give out these conditions as an alarm.
Controlled conditions on camera
and software
Standard process interface
TM-PIF-TIM
Industrial process interface
TM-PIF500V2-TIM
Interruption USB cable to camera
Interruption data cable camera - PIF
Interruption power supply - PIF
Shut-down of TIM Connect software
Crash of TIM Connect software
1
-
Fail-safe output 0 V at analog output (AO) Open contact (Fail-safe relay)/
green LED off
1) Depending on supply voltage
2) Active if AO1,2 or 3 is/ are programmed as alarm output.