Page 11
Technical Data
thermoIMAGER TIM
2. Technical Data
2.1 Functional Principle
The thermoIMAGER TIM calculates the surface temperature based on the emitted infrared energy of objects,
see Chap. 10. The two-dimensional detector (FPA - focal plain array) allows a measurement of an area and
will be shown as thermographic image using standardized palettes. The radiometric processing of the picture
data enables the user to do a comfortable detailed analysis with the software TIM Connect.
2.2 General Specifications
Model TIM 160 200 230 400 450 640
Protection class IP 67 (NEMA-4)
1
Operating temperature 0 ... 50 °C
(+32 ... +122 °F)
0 ... 70 °C
(+32 ... +158 °F)
0 ... 50 °C
(+32 ... +122 °F)
Storage temperature -40 ... 70 °C
(-40 ... +158 °F)
-40 ... 85 °C
(-40 ... +185 °F)
-40 ... 70 °C
(-40 ... +158 °F)
Relative humidity 10 ... 95 %, con condensing
Material (housing) Aluminum, anodized
Dimensions 45 x 45 x 62 - 65 mm
3
(depending on the lens)
46 x 56 x 86 - 90 mm
3
(depending on the lens)
Weight (inclusive lens) 195 g 215 g 320 g
Cable length (USB) 1 m (standard), 5 m, 10 m, 20 m
Vibration IEC 68-2-6: (sinus shaped), see Chap. 2.3
IEC 60068-2-64 (broadband noise), see Chap. 2.3
Shock IEC 68-2-27: 25 g and 50 g, see Chap. 2.3
1) Only with 5, 10, 20 m USB cable. The camera plug of USB cable (1 m) does not feature an IP 67 protection
class.
Kommentare zu diesen Handbüchern