29
Accessories page 30 - 31
Mounting bracket
Air purge collar
Rail mount adapter for controller
Water cooled housing
Interface kit
Software CompactConnect
Certificate of calibration
Model CTLC-4SF45-C3 CTLC-2SF45-C3 CTLC-6SF45-C3
Optical resolution 45:1 45:1 45:1
Temperature range
1
200°C to 1450°C
Spectral range 3.9µm 4.24µm 4.64µm
Fields of application
through flames to monitor workpieces
inside ovens, to measure inside
chemical reactors, to observe the brick
temperature in combustion chambers
CO
2
flame gases in combustion
processes, garbage burning or
processes inside chemical reactors
CO flame gases in combustion
processes, garbage burning or
processes inside chemical reactors
System accuracy
3, 4
±1% or ±1.5°C
Repeatability
3
±0.5% or ±0.5°C
Temperature resolution 0.1°C
Response time (90% signal)
2
10ms
Emissivity/gain
1
0.100 - 1.100
Transmissivity/gain
1
0.100 - 1.000
Signal processing
1
peak hold, valley hold, average; extended hold function with threshold and hysteresis
Outputs/analog
channel 1 0/4 - 20mA, 0 - 5/10V, thermocouple J, K
channel 2 sensing head temperature (-20°C to 85°C as 0 to 5V or 0 to 10V), alarm output
Alarm output 24V / 50mA (open collector)
Optional relay: 2 x 60 V DC/42 V AC
eff
; 0.4 A; optically isolated
Outputs/digital optional USB, RS232, RS485, CAN, Profibus DP, Ethernet
Output impedances
current output mA max. 500 Ω (with 8 to 36 V DC)
voltage output mV min. 100 Ω load impedance ; thermocouple 20 Ω
Inputs
programmable functional inputs for external emissivity adjustment,
ambient temperature compensation, trigger (reset of hold functions)
Cable length 3m (standard), 8m, 15m
Power supply 8 to 36VDC; max. 160mA
Laser class II (635nm), 1mW, ON/OFF via controller or software
Environmental rating IP 65 (NEMA-4)
Ambient temperature sensor: -20°C to 85°C (50°C if Laser ON) ; controller: 0°C to 85°C
Storage temperature -40°C to 85°C
Relative humidity 10 to 95%, non condensing
Vibration IEC 68-2-6: 3 G, 11 - 200Hz, any axis
Shock IEC 68-2-27: 50 G, 11ms, any axis
Weight sensor: 600g ; controller: 420g
1
adjustable via programming keys or software
2
with dynamic adaption at low signal levels
3
at ambient temperature 23 ±5°C; whichever is greater; temperature of the object >0°C
4
ε = 1, response time 1s
Cable length [3 m Standard / 8 m / 15 m]
Focus [SF45 / CF1 / CF2 / CF3 / CF4]
Spectral range [3.9 µm / 4.24 µm / 4.64 µm]
thermoMETER CTLaserCOMBUSTION
Product identification
CTLC - 4 SF45 - C3